Photon-counting detector CT without electrocardiographic gating minimizes radiation dose while maintaining high image quality in children with congenital heart disease

    This retrospective study assessed radiation dose and image quality in 37 children undergoing pediatric cardiac CT using photon-counting detector CT (PCD CT) in comparison with 37 matched patients imaged with energy-integrating detector CT (EID CT). PCD CT resulted in significantly lower radiation exposure, with a volume CT dose index (CTDIvol) of 0.20 ± 0.12 mGy versus 0.37 ± 0.17 mGy with EID CT, and a dose-length product (DLP) of 4.06 ± 3.22 mGy*cm versus 7.21 ± 4.67 mGy*cm (p < 0.001). Quantitative and qualitative image quality, including signal-to-noise ratio, contrast-to-noise ratio, and beam hardening artifacts, were comparable between the two modalities. PCD CT enabled a greater than 40% reduction in radiation dose compared with EID CT, without compromising diagnostic performance and offering a promising strategy for pediatric cardiac imaging without the need for electrocardiographic gating.